Operating Messages; Error Codes - Lovibond Tintometer MD110 Manual De Instrucciones

Ocultar thumbs Ver también para Tintometer MD110:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 218
Operating messages – Error codes
GB
Lo
btLo
Store
Date
Cal
RESULT
Time
Cal
E27 / E28 / E29
E 10 / E 11
E 20 / E 21
E23 / E24 / E25
E 22
E 72
E 73
E 74
E 75
E 80
E 81
E 82
E 83
E 84
E 85
E 86
E 87
E 88
E 89
E 90
E 91
E 92
MD110_1 12/2016

Operating messages

Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Replace batteries, no further tests possible.
Battery capacity is too low for the display backlight;
measurement is still possible.
A user calibrated method is indicated by an arrow while the test result
is displayed (see "Factory calibration reset").

Error codes

Light absorption too great. Reasons: e.g. dirty optics.
Calibration factor "out of range"
Too much light reaching the detector.
Too much light reaching the detector.
Battery capacity was too low during measurement. Change battery.
CL 6:
Factory calibration incorrect / erased
CL 6:
User calibration incorrect / erased
CL HR: Factory calibration incorrect / erased
CL HR: User calibration incorrect / erased
AL:
Factory calibration incorrect / erased
AL:
User calibration incorrect / erased
FE:
Factory calibration incorrect / erased
FE:
User calibration incorrect / erased
FE M:
Factory calibration incorrect / erased
FE M:
User calibration incorrect / erased
Cu:
Factory calibration incorrect / erased
Cu:
User calibration incorrect / erased
Zn:
Factory calibration incorrect / erased
Zn:
User calibration incorrect / erased
SO4:
Factory calibration incorrect / erased
SO4:
User calibration incorrect / erased
Mo 1: Factory calibration incorrect / erased
101

Publicidad

Tabla de contenido
loading

Tabla de contenido