ENGLISH
Important information!
Parts such as electrodes, foam pads, sponge pockets and
connector cables are subject to a certain amount of wear and
may require replacement after prolonged use. Check all parts
for damage before each application, especially the connecting
lines. If corroded, the cables are to be replaced. If the contact
plugs on the SwiSto3 become corroded, send the device in
for inspection.
The integrated rechargeable battery may only be replaced
by the manufacturer. Should the battery be defective (will
not charge, discharges too quickly), send the device in for
inspection/repair.
Warning: Modifying the SwiSto3 medical electrical device is
not allowed! In order avoid the risk of electric shock, only the
included charging plug (REF 05.19170.002) may be used to
charge the unit.
14. Storage while not in use
In order to prevent the SwiSto3 iontophoresis device from
becoming damaged, it should be stored along with its
accessories in its original packaging if not in use for a
prolonged period of time.
15. Transport and storage
When transported, the SwiSto3 iontophoresis device is to be
packaged such that it is protected from damage. If the device
is not to be used for a prolonged period of time, it should
be charged completely before being stored. The device and
accessories are to be stored in a dry location between 0 °C
and +40 °C.
16. Disposal
The non-accessible part of the SwiSto3 iontophoresis device
contains electrical and electronic components. For this reason,
at the end of its service life, the device must be brought to
an appropriate waste disposal site in accordance with local
regulations or returned to the manufacturer.
www.kawemed.com
17. Technical specifications
Classification
Protection classification
Dimensions
Nominal output current
Load resistance
Power input
Power supply
Operation mode
Weight
Safety features
Current type
Operating temperature
Storage temperature
Relative humidity
Air pressure
24
Applied part: type BF
IP 41
W 110 x H 84 x L 120 mm
Adjustable 1 ..... 25 mA
max. 1.5 kΩ
1.5 W
Li-ion rechargeable battery
Continuous operation
Charging plug: approx.
0.10 kg
iontophoresis device: approx.
0.36 kg
Electronic current limiting
device for the treatment
current, Current is cut off
when the treatment circuit is
broken, System interrupt for
when the treatment circuit is
not closed
Continual current or pulsed
current up to a max. of 25 mA
+ 10 °C to + 40 °C
0 °C to + 40 °C
30 % to 75 %
700 hPa to 1060 hPa