The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
The working position must be such as to allow one to reach the control panel and to have a good view of the LCD screen, the
suction tank and the antibacterial filter.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and
better efficiency of the overflow device.
Operation with Internal Battery
Connect the short silicon tube, with antibacterial filter, to the suction connector (VACUUM). The other tube, with one end
connected to the antibacterial filter with the other end to jar's lid connector where has been fixed the red float. When the
90% of the volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on
the jar) to the avoid liquid penetration inside the device.
Connect the long silicon tube to the other jar's lid connector (PATIENT PORT);
To start the treatment press the ON/OFF button and check its green back-light. When the device is switched on the LCD
screen activates displaying a bar-graph, graduated scale, and negative pressure value, which must be set by the operator;
The negative pressure is set with keys - and +, which are found below the LCD screen: Press one of the two keys with the
motor on. The level of the negative pressure will flash at the centre of the display.
Release the key and, after 3 seconds, the value will be operating (fixed digit on the display) and saved. The desired value
remains fixed until the operator varies the negative pressure, whilst the bar-graph moves along the semi-circumference
stopping in correspondence (graph) of the value set.
The maximum negative pressure value that can be set is: -75kPa (-0.75 bar)
The minimum negative pressure value that can be set is: -10kPa (-0.10 bar)
To suspend and / or terminate the treatment press the ON / OFF button and and pull the plug out from the power socket.
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of
the functionally vacuum) then rescrew the lid on the jar correctly.
To extract the accessories and start with cleaning.
At the end of each use place the device on its FIXING BRACKET (ref. SECURING THE NEW SUCTION UNIT)
The battery is fully charged in about 70 minutes with continuous operation.
WARNING: Before using the device, check the battery power status. Before each use proceed with charging the
battery. To maintain the device in good conditions, recharge the battery every 3 months (when not in use).
WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the
"VACUUM". A wrong connection causes immediate destruction in case of contact with sucked liquids.
WARNING: The seal hook placed on the SUPPORT is the dividing element from the 12V mains, even if the device is equipped with
the specific ON/OFF button.
Filter assembling
Suction Pump
Inlet
The device features an energy saving function which allows the device to increase its autonomy and decrease the number of revs of
the motor. In order to do so, it controls the negative pressure which is generated only when the operator suctions body fluids.
If the operator does not carry out the suction process the device lowers the number of revs of the motor, thus protecting the
duration of the internal battery.

INSTRUCTION FOR USE

NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER
FLOW DIRECTION
IN / Fluid Inside
IN
24
Jar Air Tube