LIGHT INDICATORS:
The device is provided with a light indicator (directly on the LCD screen) which allows you to view the operation of the device,
autonomy of the battery, and the charge phase in progress.
The light indications, which appear during operation, are indicated in table I.
The charge cycle starts automatically by placing the device on the specific support bracket (ref. Support Bracket and Charge
Chapter). The charge occurs only when the motor of the suction unit is off.
DISPLAY IN CHARGE: this is on with white back-light and it remains on for the entire time. It switches off if there is no external
voltage. The charge phase lasts until the device is powered and the battery is kept at its maximum charge.
The large digits display "CHA" .
TAB. I – INDICATOR LIGHTS DURING OPERATIONS
Back-light signals
Fixed white back-light and
progressive activation of the
bar-graph
Fixed white back-light and
fixed activation of the bar-
graph
Fixed green back-light
Fixed yellow back-light
Fixed red back-light
Flashing red back-light
Phase
During the Charge
phase
During the Charge
phase
During battery
operation
During battery
operation
During battery
operation
Automatic switch-off of
the device due to
discharged battery
Problem / Cause
Battery charge in progress
Charging cycle complete
Primary status / Battery
fully charged
Intermediate status /
Battery not fully charged
Low battery
Battery fully discharged
25
Solution
Wait
Remove the device from the
support plate of the ambulance
Battery operation guaranteed
Battery operation guaranteed /
start the charge cycle upon
activation of the red back-light
Start charge phase
ATTENTION: A long and
continuous beep is produced
during this signal to warn the
operator about the low charge
level of the battery.
The flashing back-light switches
on upon reactivation of the
device. Immediately start the
battery charge cycle.
GB