7-2 Removal of powder from the Powder Case, Handpiece and Nozzle
1) Remove the Powder Case from the hose.
2) Pull back the Handpiece Release Collar and
remove the handpiece.
3) Insert the Nozzle into the Ring Wrench and
loosen the Nozzle. Remove the Nozzle from the
Handpiece. (Fig. 14)
4) Fit the supplied Blower Nozzle to the coupling
and attach to the handpiece. Run idle with no
water more than 5 sec. to clear lines. (Fig. 15)
Coupling
*The Blower Nozzle in the product package depends on the joint type you purchase. Do not use a blower nozzle
other than the supplied blower nozzle.
5) Connect the Blower Nozzle to the Nozzle end
(Fig. 16) and idle with air alone. Residual
powder in the Nozzle will be discharged.
When idling, hold the Nozzle and Handpiece tight and be careful not to allow the Nozzle
to be ejected by air pressure.
CAUTION
When idling, stop water supply and idle with air alone.
7-3 Cleaning and washing
1) Remove powder from the Handpiece and
Powder Case connections and the Adjustment
Ring section.
Clean the inside of the Handpiece connection
with the supplied cleaning brush. (Fig. 17: Left)
Clean the Powder Case connections with the
supplied cleaning brush in the same way.
(Fig. 17: Right)
Blower Nozzle
(PHATELUS, MACH)
Blower Nozzle
®
®
(KaVo
MULTIflex
,
®
MULTIflex
LUX Coupling)
Handpiece
12
Ring Wrench
LOOSEN
Handpiece
Blower Nozzle
Nozzle
Powder Case
Fig. 14
Fig. 15
Fig. 16
Fig. 17