3.2 Remote control commands
104
3.2.2.27. Mode.Parameter.Presel.Cond
Mode.Parameter.Presel.DriftDisp
Mode.Parameter.Presel.DCor.Type
Mode.Parameter.Presel.DCor.Value
Parameters for SET, KFT:
.Cond: Conditioning ON/OFF
.DriftDisp: Drift display during conditioning ON/OFF.
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
3.2.2.28. Mode.Parameter.Presel.IReq
Mode.Parameter.Presel.SReq
Parameters for SET, KFT, MEAS: Automatic inquiry after the start of the de-
termination. From such an inquiry, the determination continues if the re-
quested entry/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.93)
or with &M $G, see 3.2.2.1.
$H is not possible in requests.
3.2.2.29. Mode.Parameter.Presel.LimSmplSize.Status
Mode.Parameter.Presel.LimSmplSize.LoLim
Mode.Parameter.Presel.LimSmplSize.UpLim
Limit control for the sample size.
3.2.2.30. Mode.Parameter.Presel.ActPuls
Output of a pulse on the I/O line "Activate", see page 174.
3.2.2.31. Mode.Parameter.CtrlPara.EP
Mode.Parameter.CtrlPara.UnitEp
Parameters for KFT: Setting of the EP, resp. Control point as a pH value, in
mV (with U and Ipol) or µA (with Upol). The corresponding unit can be read
with .UnitEP.
With KFT the standard value for Ipol is 250 mV, for Upol 25 µA.
auto, man., OFF
id1, id1&2, all, OFF
value, unit, all, OFF
0.0...999 999
0.0...999 999
for SET, KFT: first, all, cond., OFF
pH:
0...±20.00, OFF
U, Ipol: 0...±2000, OFF
Upol: 0...±200.0, OFF
ON, OFF
ON, OFF
0.0...99.9
ON, OFF
ON, OFF
read only
758 KFD Titrino