Connect the short silicon tube, with antibacterial filter, to the suction connector.
•
The other tube already connected to the filter has to be connected to the "VACUUM" jar outlet, where has been fixed the
red float (security float).
When the 90% of the volume of the jar is reached there is the activation of the security float (the float close the aspiration
connector on the jar) to avoid liquid penetration inside the device.
The device must be used on a plan of horizontal operation.
WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the
"VACUUM". A wrong connection causes immediate destruction in case of contact with sucked liquids.
•
Connect the longer silicone tube to the "PATIENT" jar outlet
•
Connect the other end of the long silicon tube to the probe plastic connector then connect the suction probe to it.
•
Connect the power cord to the device then connect the plug to the electrical mains supply.
•
Push switch on position I to start suction.
•
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of
the functionality vacuum) then rescrew the lid on the jar correctly.
•
During operation the jar has to be in vertical position to avoid overflow valve to cut off aspiration. Should this happen,
switch off the device and disconnect the tube from the jar cover (from "VACUUM" outlet).
•
Once finished push switch on O position and unplug.
NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER
MAKE SURE THAT CHILDREN AND/OR MENTALLY ILL PEOPLE DO NOT USE THE DEVICE WITHOUT ADULT
ALWAYS PLACE THE DEVICE IN POSITIONS FOR EASY DISCONNECTION.