Safety Information

XW
To prevent possible electrical shock or personal
injury, follow these guidelines:
Before and after testing, confirm that the Tester
does not indicate the presence of a hazardous
voltage, see Figure 3. If the Tester beeps
continuously and a hazardous voltage is shown
on the display, remove power from the circuit
under test or allow the installation capacitance to
fully discharge.
Use the Tester only as specified in this manual.
Otherwise the protection provided by the Tester
can be compromised.
Connect the common test lead before the live test
lead and remove the live test lead before the
common test lead.
Do not disconnect the test leads before a test has
been completed and the test voltage at the
terminals has returned to zero. This ensures that
any charged capacitance is fully discharged.
Disconnect circuit power and discharge all high-
voltage capacitors before you measure resistance
or capacitance.
Do not work alone or around explosive gas, vapor
or dust.
Do not use the Tester in a damp or wet
environment.
Examine the test leads for damaged insulation or
exposed metal. Check test lead continuity.
Replace damaged leads. Do not use the Tester if it
looks damaged.
Be careful around voltages >30 V ac rms, 42 V ac
peak, or 60 V dc. Such voltages pose a shock
hazard.
Keep fingers behind the finger guards on the
probes.
Warning: Read before Tester use.
Insulation Tester
Safety Information
3