The actual test voltage can be up to 10 % higher than the selected test
voltage.

Program a Test Voltage

To set a test voltage in between the preset test voltages, proceed as follows:
1. With the Tester turned on, push  to select TEST VOLTAGE.
2. Push  or  to scroll through the preset test voltage options (250 V,
500 V, 1000 V, 2500 V, 5000 V, and 10,000 V). Select the voltage closest
to the level required.
3. The selected test voltage appears in the upper-right of the display.
4. Push . TVxxxxV appears flashing in the lower-left of the display.
5. Push  or  to increment and decrement the voltage. When the
correct voltage level shows, do not push . Doing so will return the test
voltage to the next lowest last preset voltage selection. Instead, push 
to go to the function menu.
The test voltage can be up to 10 % higher than the test voltage you
select.
Note
Note
Insulation Tester
Measurements
Test Voltage Display
GHH05.EPS
17