4.2. Control commands
4.2.3 G
'GO' triggers dosing in all modes.
In mode DOS dosing goes on until a stop command is received.
4.2.4 S
'STOP' terminates dosing in modes DOS, DIS R and DIS C (not filling).
4.2.5 F
'FILL' triggers filling of the burette in all modes. May also serve as emergency stop.
If the exchange unit is already filled, filling is not executed.
4.2.6 C
'CLEAR' resets volume display to 0.000 ml.
4.2.7 I
Command 'I' (information) makes the Dosimat send information bytes terminated by 'CR'
+ 'LF'. These information bytes contain the information READY, CODE, LOCK, LIMIT etc.
Command 'I' is also active at "REM OFF".
Information byte 1:
Cylinder code:
Bit
0
1
2
36
GO, not live, 1-byte command
STOP, live, 1-byte command
FILL, live, 1-byte command
CLEAR, not live, 1-byte command
Information, live, 1-byte command
Bit
Function
0
Cylinder code
1
Cylinder code
2
Cylinder code
3
1 =no exch. unit
4
1 = New exchange unit
5
1 = Ready
6
1 =V-LIM reached
1 ml
5 ml
0
1
1
0
1
0
10 ml
20 ml
1
1
1
0
1
1
50 ml
1
1
0
765 Dosimat