4.2. Control commands
If a value is entered which is too high or too low, resp. , the Dosimat corrects it
automatically to volume
4.2.7).
4.2.32 VLI OFF
Security volume control is switched off.
In modes DOS, DIS C, and PULSE only, in other modes the command is not accepted
and bit 0 of information byte 2 is set to 1.
The value of V-LIM is overwritten.
4.2.33 AFI ON
Function 'auto fill' is switched on.
4.2.34 AFI OFF
Function 'auto fill' is switched off.
4.2.35 QDI
Request to send content of display via RS232 interface.
For example: "DIS C
4.2.36 QVO
Request to send the volume value via RS232 interface.
Do not query the volume when result calculation in Mode DOS is active!
For example: " 13.457 'CR' 'LF'"
4.2.37 QPO
Request to send the content of the position counter via RS232 interface.
The position counter (16 bit binary) gives the actual position of the piston.
0 = Initial position (filled)
10000 = Final position (empty)
42
Cylinder
Volume
ml
ml/min
1
0.001
5
0.001
10
0.001
20
0.002
50
0.005
or volume
resp. and sets bit 1 of information byte 2 to 1 (see
max
min,
Security volume off, not live
Auto fill on, live
Auto fill off, live
Query display, live
13.457 ML 'CR' 'LF'"
Query volume value, live
Query position of piston, live
Volume
min
max
ml/min
999.999
999.999
999.999
999.999
999.999