APPendIx II
System Alarm Prompt
PROMPT
"XX TOO HIGH"
"XX TOO LOW"
XX represents the value of parameter such as HR, ST1, ST2, RR, SpO
"ECG WEAK SIGNAL"
"NO PULSE"
"RESP APNEA"
"CO2 APNEA"
"ASYSTOLE"
"VFIB/VTAC"
"COUPLET"
"BIGEMINY"
"TRIGEMINY"
"R ON T"
"PVC"
"TACHY"
" BRADY"
"VT>2"
"MISSED BEATS"
CAUSE
XX value exceeds the higher alarm
limit.
XX value is below the lower alarm
limit.
The ECG signal of the patient is too
small so that the system can not per-
form ECG analysis.
The pulse signal of the patient is too
small so that the system can not per-
form pulse analysis.
The respiration signal of the patient is
too small so that the system cannot
perform RESP analysis.
The respiration signal of the patient is
too small so that the system cannot
perform RESP analysis.
Patient suffers from Arr. Of ASYSTO-
LE.
Patient suffers from Arr. of VFIB/VTAC.
Patient suffers from Arr. of COUPLET.
Patient suffers from Arr. Of BIGEMINY.
Patient suffers from Arr. of TRIGE-
MINY.
Patient suffers from Arr. of R ON T.
Patient suffers from Arr. of PVC.
Patient suffers from TACHY.
Patient suffers from BRADY.
Patient suffers from Arr. of VT>2.
Patient suffers from Arr. of MISSED
BEATS.
MEASURE
Check if the alarm limits are appropriate
and the current situation of the patient.
, IBP, NIBP, etc in the system.
2
Check if the electrodes and lead wires
are connected correctly and the current
situation of the patient.
Check the connection of the sensor and
the current situation of the patient.
Check the connection of the linking wire
and the current situation of the patient.
Check the connection of CO2 sensor and
the current situation of the patient.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
Check the current situation of the patient.
Check the connection of the electrodes
and lead wires.
104