Technical specifications
90W power supply unit
AC input range:
DC output:
Typical power consumption:
Peak power consumption:
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Electromagnetic compatibility
The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC
60601-1-2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication
devices are kept at least 1 m away from the device.
For further details see the Guidance and manufacturer's declaration – electromagnetic emissions and immunity section.
IEC 60601-1:2005 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensors:
Flow sensor:
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
O (30 hPa) for more than 6 sec or 40 cm H
2
Sound
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
Standard:
SlimLine or Standard and humidification:
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine:
Standard:
SlimLine or Standard and humidification:
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Alarm volume settings
Low (nominal 54 dBA), Medium (nominal 60 dBA), High (nominal 73 dBA)
100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
24V
3.75A
53W (57VA)
104W (108VA)
+41°F to +95°F (+5°C to +35°C)
Note: The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions
(104°F/40°C) the device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to
738 hPa
-4°F to +140°F (-20°C to +60°C)
5 to 95% relative humidity, non-condensing
Internally located at device outlet, analog gauge pressure
type, -5 to +45 cm H
2
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
O (40 hPa) for more than 1 sec.
2
25 dBA with uncertainty of 2 dBA
25 dBA with uncertainty of 2 dBA
27 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
35 dBA with uncertainty of 2 dBA
O (-5 to +45 hPa)
English
19