8-2 Cleaning
Manual Cleaning (External Surface)
1) Remove dirt and debris on the handpiece surface using a
brush under running water (at 38°C or less). Do not use a
wire brush.
2) Wipe off the moisture on the surface using absorbent cloth.
CAUTION
• If water comes inside the handpiece, blow air into the rear of
the handpiece using an air syringe and remove the remaining
water.
12
Automatic Cleaning (External and Internal Surfaces)
(Only for FX15m / FX23m / FX25m / FX75m)
Use a thermo-disinfector in accordance with ISO 15883
(EN ISO 15883).
Refer to the thermo-disinfector manual.
CAUTION
• After washing with thermo-disinfector and prior to
lubrication, dry the product until all internal moisture
is thoroughly removed. Thermo-disinfector moisture
remaining inside the product could reduce the effect
of lubrication and could cause corrosion inside of the
product.