NOTICE
• NSK recommends the use of "Spray Mist Absorber"
(Y900084) to prevent oil mist expelling out of the
handpiece head.
NSK automatic handpiece cleaning and lubrication
system
When using NSK automatic handpiece cleaning and
lubrication system, refer to the system instructions.
WARNING
• If blood infiltrates inside a handpiece, an
automatic handpiece cleaning and lubrication
system may not totally clean the internal
handpiece components, which may lead to
internal blood coagulation. Coagulated blood
inside the handpiece may cause handpiece
failure and overheating, causing burn injuries.
To avoid the risk, NSK recommends to
immediately lubricate those handpieces with
PANA SPRAY Plus / PANA SPRAY.
15