Code 22 screen
There is no disc in the I-neb, no
treatments left on the disc, disc has
expired, I-neb failed to read the disc, or
electrical equipment is located nearby
and is interfering with the I-neb.
Solution: Stand I-neb upright on a flat surface.
Check that the disc is correctly in place.
Remove and reinsert or replace the disc.
Turn off and move away from any electrical
equipment nearby. Reset I-neb by pushing the
ON button and holding it in. Wait for the start
up screen to appear. Pick up the I-neb and
start your treatment. (Code 22 will remain on
the screen for 10 minutes or until the ON button is
pressed for 5 seconds as described above).
Code 33 screen
Your I-neb needs maintenance. The I-neb
can continue to be used for up to 30 days
after the appearance of code 33 is seen
on the screen. Pick up the I-neb and start
treatment as normal.
Solution: Contact the patient support to
arrange return of your I-neb.
Additional information
12035_RRDD_InebPH_InstESUS_1056844_Issue1_v10.indd 19
Code 44 screen
Memory failure.
Solution: Reset by pressing the ON button
and waiting for the start up screen to appear.
Pick up I-neb and start your treatment. (Code
44 will remain on the screen for 10 minutes or
until the ON button is pressed for 5 seconds).
Code 55 screen
There is no medication in the
medication chamber at the beginning of
treatment.
Solution: Stand the I-neb upright on a flat
surface. Add medication to the medication
chamber. Reset by pressing the ON button
and waiting for start up screen to appear. Pick
up the I-neb and start your treatment. (Code
55 will remain on the screen for 10 minutes or
until the ON button is pressed for 5 seconds).