Cardiopulmonary support (CPR) should be initiated immediately per hospital
protocol if indicated for any patient supported with the Impella RP
Catheter. When
®
initiating CPR, reduce the Impella RP
Catheter flow rate. When cardiac function
®
has been restored, return flow rate to the previous level and assess the placement
signal on the controller.
During defibrillation, do NOT touch the Impella RP
Catheter, cables, or Automated
®
Impella
Controller.
®
Avoid overinserting the Impella RP
Catheter and possibly impinging the catheter tip
®
against the walls of the vasculature, atrium, or ventricle.
Torquing the catheter should be monitored carefully using fluoroscopy.
Do NOT advance or withdraw the Impella RP
Catheter against resistance without
®
using fluoroscopy to determine the cause of the resistance. Doing so could result
in separation of the catheter or guidewire tip, damage to the catheter or vessel, or
perforation.