Specification
Ground Loop Resistance (Autorange):
Operating temp.:
Range
0.025 - 0.250 Ω
0.250 - 9.999 Ω
10.00 - 99.99 Ω
100.0 - 199.9 Ω
200.0 - 400.0 Ω
400.0 - 600.0 Ω
600.0 - 1500 Ω
(1)
Loop resistance with no inductance, external field < 50 A/m, external electrical field
< 1 V/m, conductor centered.
Continuity
Operating temp.:
Beep if resistance < approximately 40.00 Ω
Ground/Leakage Current mA~
Operating temp.:
(Autorange, 50/60 Hz, True RMS, crest factor CF< 3.0)
Range
0.200 - 1.000 mA
1.00 - 10.00 mA
10.0 - 100.0 mA
100 - 1000 mA
Ground/Leakage Current A~
Operating temp.:
Range
0.20 - 15.00 A
15.00 - 30.00 A
Max. Current Protection: 100A continuous, 200A (< 60 sec.) 50/60Hz.
Current Overload Protection: 100A Continuous, 200A ( < 60 seconds) 50/60Hz.
20
0 °C ... +50 °C (+24 °F ... +122 °F)
0 °C ... +50 °C (+24 °F ... +122 °F)
-10 °C ... +50 °C (+14 °F ... +122 °F)
Resolution
0.001 mA
0.01 mA
0.1 mA
(50/60 Hz, True RMS, crest factor CF< 3.0)
-10 °C ... +50 °C (+14 °F ... +122 °F)
Resolution
0.01 A
0.01 A
Resolution
0.002 Ω
0.02 Ω
0.04 Ω
0.4 Ω
2 Ω
5 Ω
20 Ω
1 mA
Operating error (1)
±1.5%rdg ± 0.05 Ω
±1.5%rdg ± 0.1 Ω
±2.0%rdg ± 0.3 Ω
±3.0%rdg ± 1.0 Ω
±5.0%rdg ± 5 Ω
±10%rdg ± 10 Ω
±20%
Operating error
±2.5%rdg ± 0.05 mA
±2.5%rdg ± 0.05 mA
±2.0%rdg ± 0.3 mA
±2.0%rdg ± 3 mA
Operating error
±2.0%rdg ± 0.03 A
±3.0%rdg ± 0.03 A
HEME GEO 30