5. Press the slider on the sensor head until the UNLOCK indication disappears,
and hold it until LOCK appears. Check that the opening lever is firmly locked
and the clamp is securely closed.
Correct measurements cannot be performed unless the clamp is securely closed
and the slider is pressed until LOCK is displayed.
The current consumption of clamp-on probes depends on the current to be
measured. Confirm that the total current consumption of the probes does not
exceed the rated output current of the R&S RT-ZA13 probe power supply.
Fig. 3-10: Current consumption* vs. current to be measured (typical)
*The sum total of a positive and negative current consumption
User Manual 1409.7808.02 ─ 01
Putting into Operation
Connecting the Probe to the DUT