Mount or replace the mercury lamp
Before replacing the bulb, please set the fluo-
rescence power supply to (OFF) position and
unplug the power cable. Be sure that the lamp
has completely cooled down.
Loosen the lock-screw
and take off the bulb holder
lock screw
and take off the plastic pole
(Fig.25)
The HBO lamp has 2 poles of different dimen-
sions, so that it fits in only one direction: insert
the anode pole (bigger one) into the fixed hold-
er
and the other pole in the flexible holder.
Then tighten the screw. (Fig.26)
Mount the lamp door back in place and tighten
the door locking screw.
(Fig.24) completely
. Loosen the
.
Page 11
Fig.24
Fig.25
Fig.26