Chuck cleaning
Clean Push Button chuck once a week.
1) Mount the arrow-head spray nozzle tip into the spray can port.
2) Lubricate the chuck directly through the bur insertion hole. (Fig. 11)
3) Lubricate handpiece by using NSK PANA SPRAY Plus (Fig. 10) or Care3 Plus.
CAUTION
• If the chuck is not regularly cleaned the chuck grip may be weakened and the bur may be
accidentally released while in operation.
7-5 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur after each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.