5. Operation
5.1 Before switching on
•
If the device has been stored in a cold en-
vironment (below +10 °C) before being
switched on, put it in a warm room and
wait approximately 30 minutes before
switching it on to allow the OXYMAT 3 to
come to room temperature.
•
Unscrew humidifier glass
humidifier from dosage monitor
glass with distilled (aqua dest. from your
pharmacy), demineralised or boiled wa-
ter up to the mark. Then screw humidifier
glass on again.
•
If you use a sterile water system, fit this in
accordance with the appropriate installa-
tion instructions.
5.2 Switching on and off
1. Switch on the OXYMAT 3 using On/Off
knob
. A self-test will then run.
2
– Red fault indicator
dicator
and green power indicator
19
immediately come on briefly.
20
– Green power indicator
es for the duration of the self-test (ap-
prox. 1 minute).
– During the self-test you will hear the
compressor stopping and system pres-
sure being released after approx.
30 seconds (hissing noise). This is an
internal leak test taking place. The
compressor then starts up again.
– If the device is functioning perfectly,
only green power indicator
be on after the self-test.