Range Selection: DIP Switches
View on DIP switches S8 ... S1 and LED (top view)
Green
Red
Measured variable
Memosens pH/ORP
Memosens OXY
Memosens COND
Memosens CONDI / SE 670
Table: x means switch = on, gray: default setting

Measuring range

pH/ORP
pH 0 ... 14
pH 0 ... 7
pH 2 ... 12
pH 4 .. 12
pH 5 ... 10
pH 7 ... 14
–1200 ... 1200 mV
– 500 ... 500 mV
Table: x means switch = on, gray: default setting
OXY
0 ... 120 % (AIR)
0 ... 10 % (AIR)
0 ... 50 mg/l
0 .. 10 mg/l
0 ... 1000 µg/l
0 ... 100 µg/l
0 ... 25 % (GAS)
0 ... 1 % (GAS)
Cond/CondI
0 ... 2000 mS/cm
(uncompensated)
0 ... 100 mS/cm
(uncompensated)
0 ... 1000 µS/cm
(uncompensated)
0 ... 10 µS/cm
(uncompensated)
0 ... 2000 mS/cm
(TC 2.1 %/K)
0 ... 100 mS/cm
(TC 2.1 %/K)
0 ... 1000 µS/cm
(TC EN 27888)
0 ... 10 µS/cm
(TC ultrapure water)
S5
S4 S3
x
x
x
x
x
x
x
x
x
x
x
x