Error Messages On The Display - Bühler technologies TC-Standart Guía Rápida

Refrigerador termoeléctrico de gases
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 25
TC-Standard (+)
6 Maintenance
Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator's manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.
7 Service and Repair
Please refer to the original operator's manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.

7.1 Error messages on the display

If an error occurs, the display will read "Err". Press the "
button to show the error number(s).
Error messages will appear until the unit has been restarted
or the error is cleared using the "Func" button. It can only be
cleared if the cause for the error has been corrected.
Causes / Action: The following is a list of the most common
causes and actions for the respective error. If the actions listed
do not resolve the problem, please contact Service.
Problem / Malfunc-
Possible cause
tion
No display
• No voltage
• Loose connect-
• Display defect-
D1.02
(The software ver-
sion for the display
will appear).
(permanent)
• Not communic-
Error
• An error has oc-
Error 01
• Controller mal-
Error 03 • µC fault / MCP2
Error 04 • EEPROM error
Error 22
• Moisture de-
BX440017 ◦ 03/2017
Action
• Check the
supply cable
ing cable
• Check fuse
• Check con-
ive
nections
• Check con-
nections
ating with the
controller
• Read the er-
curred
ror number
as described
above
• Clear error
function
(temporary
fault)
• Disconnect
from power
for approx. 5 s
• Contact ser-
vice
• Contact ser-
vice
• Contact ser-
vice
• Check mois-
tector 1 cable
ture detector
break
line
• Check mois-
ture detector
Error 32
Error 40 • General error
Error 41
Error 42
"
Error 43
Error 50
Error 51
Error 52
Error 53
• Moisture de-
• Check mois-
tector 2 cable
ture detector
break
line
• Check mois-
ture detector
• Sensor pos-
temperature
sibly defective
sensor 1
• Low temperat-
• Check tem-
ure / short-cir-
perature
cuit temperat-
sensor con-
ure sensor 1
nection
• Excess temper-
• Check tem-
ature / short-
perature
circuit temper-
sensor con-
ature sensor 1
nection
• Measurement
• Check tem-
fluctuation
perature
temperature
sensor con-
sensor 1
nection
• µC fault
• General error
• Sensor pos-
temperature
sibly defective
sensor 2
• Low temperat-
• Check tem-
ure / short-cir-
perature
cuit temperat-
sensor con-
ure sensor 2
nection
• Excess temper-
• Check tem-
ature / short-
perature
circuit temper-
sensor con-
ature sensor 2
nection
• Measurement
• Check tem-
fluctuation
perature
temperature
sensor con-
sensor 2
nection
• µC fault
Bühler Technologies GmbH