OPERATING INSTRUCTIONS (CONT)
Weighing a Patient
1. Press ON/ZERO to power on the scale. If the scale is already on, press ON/ZERO to zero the
scale. Note: ON/ZERO should be pressed before each weighing.
2. When "0" appear on the display, place baby on tray.
3. Allow scale to lock onto the weight. The scale will beep and display "PRIOR WT" when the
weight is locked.
4. Remove baby from tray. Measurement value remains on the display until it is cleared.
5. To clear the displayed weight and ready the scale for the next weighing, press ON/ZERO.
Tare Function
When using this scale, the weight of an object, such as a blanket, on the scale with the baby being
weighed can be subtracted from the total weight to determine the net weight of the baby alone. The
Tare function automatically performs this subtraction, avoiding the need for manual calculations.
1. Turn the scale on and wait for the display to read "0".
2. Place the object (i.e. blanket) on the tray. The weight of the object will be displayed.
3. Press TARE. "NET WT" and "0.0" will be displayed.
If the object is removed from the platform, its weight will be displayed as a negative value.
4. Place the baby on the tray with the object. The readout will now display the baby's net
weight.
5. Press ON/ZERO to return the scale back to normal weighing operation.
Weightlock Function
When using this scale, the weight of any object over 512 grams will be locked on the display even
after the baby has been removed from the tray. This allows the caregiver to tend to the needs of the
baby first before recording the weight. This value can be recalled even after the scale has turned off.
When the baby's weight is locked, "PRIOR WT" is displayed. The weight will remain displayed even
after the baby is removed from the tray. No further weighings can take place until the weight is
cleared by pressing ON/ZERO or REWEIGH/RECALL .
P/N UM2210KL 2210KL4 Rev20190312_Web
TARE
Tare button
gram
Net weight display
gram
Weightlock display
REWEIGH
RECALL
Reweigh/Recall button
12
NET WT
PRIOR WT.