NSK Varios Combi Pro Manual De Instrucciones página 18

Ocultar thumbs Ver también para Varios Combi Pro:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 122
8-2 Powder Preparation
1) Select the Chamber and powder to be used with this Mode. Fill the Chamber with
powder up to the "MAX" line (approximately 80g of FLASH pearl or approximately
50g of Perio-Mate powder) (Fig. 24).
Mode
Prophy Mode
For Prophy use
Perio Mode
For Perio use
2) Close the Chamber Cap securely until it fits the Chamber completely and insert the
Chamber straight to the Control Unit (Fig. 25). The type of inserted Chamber is
displayed on the LCD Panel as one of the following icons (Fig. 26).
• Use the designated powder only. Using other powders may lead to product failure, damage to the tooth or may cause
emphysema. In addition, it may also cause performance reduction of the product. We cannot accept any liability for failures
WARNING
caused by using other powders.
• Securely attach the Chamber Cap to prevent the chamber cap
from being blown off, which may lead to an injury (Fig.27).
CAUTION
• Before inserting the Chamber into Control Unit, make sure the
O-Rings of the Chamber Base Connector are not damaged or
torn (Fig.25). Damaged or torn O-Rings could lead to poor
operation or malfunction.
• Insert the Chamber into the Control Unit straight. Inserting at an angle could damage the O-Rings, leading to malfunction.
• If too much cleaning powder is put into the Chamber, the correct amount of powder may not be sprayed immediately after starting use.
• Do not use powder that has been stored in an open package or leftover powder. Using such powder may cause clogging of the
Nozzle. Do not mix leftover powder with new powder.
• Before inserting the Chamber, check the Chamber Base Connector for water, etc., and wipe it clean if it is wet. Failing to do so
may cause powder clogging.
• Do not use discolored powder.
• Always securely close the powder container lid. Failure to do so will attract moisture which may cause clogging.
• If powder adheres to the Chamber and/or the Chamber Cap, clean it immediately or the Cap will become difficult to close.
• Do not invert the Chamber while it contains powder inside. Inverting the Chamber could lead to powder clogging.
• As a safety feature, this product does not operate if the Foot Control is depressed when the Chamber is mounted incorrectly. An
alarm sounds heard and the tooth icon and "POWDER" flash alternately. Mount the Chamber correctly.
• The Perio-Mate Powder (option) scatters easily because it is made of very fine particles. Fill the Chamber slowly to prevent the
powder from scattering, and be careful not to swallow the powder or allow it to get into eyes.
• FLASH pearl is almost insoluble in water. To avoid clogging of the pipes by using FLASH pearl, periodical cleaning and
maintenance of the vacuum hose, saliva ejector and separator are necessary.
Chamber
Powder
FLASH pearl
Perio-Mate powder
Perio Mode
Fig. 25
17
"MAX" line
Fig. 24
Prophy Mode
Fig. 26
Fig. 27

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido