Intended Use; Scope Of Delivery; Electrical Voltage; Potentially Explosive Atmosphere - Bühler technologies EMIDos Guía Rápida

Unidad de dosificación fija
Ocultar thumbs Ver también para EMIDos:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 28
EMIDos
1 Introduction
This quick guide will assist you in starting up the unit. Follow
the safety notices or injury to health or property damage may
occur. Carefully read the original operating instructions in-
cluding information on maintenance and troubleshooting
prior to startup. These are located on the included CD and on-
line at
www.buehler-technologies.com
Please direct any questions to:
Bühler Technologies GmbH
Harkortstraße 29
40880 Ratingen
Germany
Tel.: +49 (0) 21 02 / 49 89-0
Fax: +49 (0) 21 02 / 49 89-20
These operating instructions are a part of the equipment. The
manufacturer reserves the right to change performance-, spe-
cification- or technical data without prior notice. Please keep
these instructions for future reference.

1.1 Intended use

This unit is intended for industrial use in gas analysis systems.
It is a component for conditioning sample gas to reduce the
loss of water-soluble gas components.
Please note the specifications in the data sheet on the specific
intended use, existing material combinations, as well as pres-
sure and temperature limits.

1.2 Scope of delivery

• Stainless steel container bracket for acid container with
fastening strap,
• Acid container
• Modified container cover,
• Metering pump with PFA dosing line,
• Connection packs,
• Capacitive sensor (optional),
• Collection pan (optional),
• Product documentation
2 Safety instructions
The equipment must be installed by a professional familiar
with the safety requirements and risks.
Be sure to observe the safety regulations and generally applic-
able rules of technology relevant for the installation site. Pre-
vent malfunctions and avoid personal injuries and property
damage.
The operator of the system must ensure:
• Safety notices and operating instructions are available
and observed,
• The respective national accident prevention regulations
are observed,
• The permissible data and operational conditions are
maintained,
• Safety guards are used and mandatory maintenance is
performed,
• Legal regulations are observed during disposal,
• compliance with national installation regulations.
BX450031 ◦ 08/2020
• The current and voltage supply for the aggregate has a
(mains) separator with adequate switching capacity. Na-
tional requirements must be observed.
DANGER

Electrical voltage

Electrocution hazard.
a) Disconnect the device from power supply.
b) Make sure that the equipment cannot be reconnected to
mains unintentionally.
c) The device must be opened by trained staff only.
d) Regard correct mains voltage.
DANGER
Toxic, corrosive medium
Medium can be harmful.
a) If necessary, ensure the medium is discharged safely.
b) Always switch off the dosing pump to stop the medium
flow before performing maintenance and repairs.
c) Use protection when handling toxic/acidic mediums.
Wear suitable protective equipment.
d) Observe the safety notices in the safety data sheets of the
acid manufacturers.
e) Observe the relevant regulations on handling corrosive or
toxic mediums.
DANGER

Potentially explosive atmosphere

Explosion hazard if used in hazardous areas.
The device is not suitable for operation in hazardous areas
with potentially explosive atmospheres.
Do not expose the device to combustible or explosive gas mix-
tures.
3 Transport and storage
Only transport the product inside the original packaging or a
suitable alternative.
The equipment must be protected from moisture and heat
when not in use. It must be stored in a covered, dry, dust-free
room at room temperature.
The container may only be transported or stored with the ori-
ginal cover tightly sealed.

4 Installation and connection

4.1 Installation site requirements

This device is solely intended for use in enclosed areas. Ensure
accessibility during operation and maintenance.
Mount the container bracket so there is sufficient room for a
collection pan below the dosing unit. This collection pan must
be made from acid-resistant material and hold at least 10 L of
liquid (please note the attached safety data sheet regarding
storage). The operator must determine the need for a collec-
tion pan for discharged media as part of the risk assessment.
Bühler Technologies GmbH
11

Publicidad

Tabla de contenido
loading

Tabla de contenido