Maintenance; Cleaning The Compressor's Outer Case; Cleaning The Nebulizer Cup - Graham Field JOHN BUNN Neb-u-Tyke Bella the Butterfly Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MAInTenAnCe

Cleaning the Compressor's outer Case

WARNINg: electrical shock hazard. Do not remove the
outer case of the device. All disassembly and maintenance
of the device must be performed by a qualified service
technician. Refer service to qualified service personnel.
WARNINg: This device does not require oil. Do not
attempt to lubricate any internal parts.
WARNINg: Unplug device before cleaning. Do not
submerge in water for cleaning.
Wipe the compressor's outer case with a clean, damp
cloth weekly to keep it dust-free. Do not use any powdered
cleaners or soap.

Cleaning the nebulizer Cup

All parts of the disposable nebulizer, except tubing, should
be thoroughly cleaned after each use. If your physician
specifies a specific cleaning procedure, follow that
procedure.
WARNINg: To prevent possible risk of infection from
contaminated medication, clean the nebulizer after each
treatment.
1. After each use, disassemble the nebulizer cup and
mouthpiece completely, wash all parts in warm soapy
water, and rinse thoroughly. Allow to air dry. Store in a
clean, dry location.
2. Once daily, after washing as above, soak nebulizer
components in a fresh solution consisting of one part
warm water and one part white vinegar for 30 minutes
followed by a warm water rinse. Allow to air dry. Store
in a clean, dry location.
Bella the Butterfly Nebulizer Compressor
13

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Jb0112-071

Tabla de contenido