6607290.book Seite 34 Montag, 5. September 2016 11:53 11
5 Operation
34
Operating Instructions Midwest Stylus
NOTICE! Use only burs and diamond polishers that are sharp and
Use only undamaged burs
undamaged. Use clean burs and diamond polishers to avoid dirt in the
clamping system.
CAUTION! A loose or partially removed bur instrument can detach
Drill Fixation _ Handpiece
itself from the head or break off. This may cause injury! Therefore, use
the handpiece only if the bur instrument is inserted at least 10 mm and
is clamped securely in place.
Inserting the friction grip bur
Attach and Remove FG _Turbine
✔ The handpiece is at a standstill.
1. Press the button and insert the bur up to the stop.
2. Pull on the bur to check that it is firmly attached.
Removing the friction grip bur
✔ The bur must not be moving.
➢ Press the button and remove the bur.
®
Plus Air-Driven-High-Speed Handpiece
6607290 D3673.201.01.01.02