6607290.book Seite 42 Montag, 5. September 2016 11:53 11
7 Conditioning
42
Operating Instructions Midwest Stylus
1. Brush the handpiece under running water (< 38 °C, < 100 °F, at
least drinking water quality) and good lighting (min. 500 lux) and
color rendering index (min. 80 Ra) until no more dirt can be seen, for
at least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and
adapters according to manufacturer's instructions.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds.
5. Conduct thermal disinfection or unwrapped steam sterilization.
6. Maintain mechanical parts manually [ → 43].
7. Maintain the push button chuck manually [ → 45].
8. Pack the handpiece in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
9. Sterilize the adapter [ → 47].
®
Plus Air-Driven-High-Speed Handpiece
6607290 D3673.201.01.01.02