Par 07
Ordinary and extraordinary maintenance
The machine does not need any particular maintenance apart from the periodic cleaning of the heating unit
which is necessary to avoid the deposit of substances inside the seat of the biological indicators. Use a wet
cloth for cleaning. Do not use aggressive cleansers.
DO NOT DO MAINTENANCE OR INSIDE REPAIR TO THE DEVICE.
DISMANTLE THE DEVICE MEANS BREAKING IT
Par 08
Analysis and research of failures
N° Difetto
Pushing start
1
button see-test
does not work
With led switch on
see-test does not
2
reach the selected
temperature
Led switch off during
3
the test
33
Causa
There is no power
supply
Anomalies in the
electronic card or
in the thermostat
Electrical energy
interruption
Inside failure
Intervento
Control plug fit in correctly
Contact Tecno-gaz s.p.a. for technical assistance
Push start button, if led don't switch on, control cable and
power plug. If led switch on, remove tubes and keep
device switch on for about 30 minutes. If device don't
switch off automatically, means that the previous
stopping was due to a temporary lack of electrical
energy. Otherwise if device switch off, means that the
heating unit work over the selected temperature and the
safety device become effective. In this case contact
Tecno-gaz s.p.a.
Contact Tecno-gaz s.p.a. for technical assistance
ENGLISH
DSTF080_REV5