Gima ECG 300G Manual De Instrucciones página 152

Ocultar thumbs Ver también para ECG 300G:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

The biological ECGs are fed into the device in form of digital signals, then the
measurement value can be obtained by calculation.
Test condition:
a) without NOISE
b)with 25uV high frequency
c) with 50uV peak to valley 50Hz/60Hz sinusoidal line frequency NOISE
d) with 1mV peak to valley 0.3Hz sinusoidal base-line NOISE
For each NOISE level above, the differences of measurements between the NOISE-free
ECGs and the ECGs with NOISE shall be determined. The two largest deviations from the mean
shall be estimated before calculation of mean and standard deviation of differences.
5.2 Verification of interpretation function
5.2.1 Verification process
5.2.1.1 CSE diagnostic database
Read initial DCD files
Frequency conversion
Voltage conversion
Get ECG data files
Automated interpretation item
Compare automated interpretation with expert diagnosis
Summarize comparison results, draw conclusions
Start
End
59

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

33221

Tabla de contenido