The loupes are correctly adjusted when you see a clearly-defined, round, shadow-free field of view.
Please consult the Instruction leaflet for HEINE
The illuminated field of view can be selected by means of the adjusting ring (4).
The adjusting ring can be fitted with a sterilisable sleeve.
fitting HEINE binocular loupes and a video camera: Use the appropriate adaptor, which is available as an
accessory.
Supporting the fiber optics cable
The F.O. cable includes adjustable clips which are used to attach the cable to a gown or similar.
HyGIENIc cLEANING / PROcESSING
Use a micro-fiber cloth to clean the lens. The instrument housing can be cleaned with a soft cloth moistened with
alcohol or a non-aggressive disinfectant solution. Do not soak or spray with disinfectant.
The adjuster sleeve can be autoclaved at 134°C for 5 minutes.
Read the instructions for cleaning the HEINE
Maintenance
This product does not require regular maintenance.
Service
This product does not require regular service.
disposal
The product must be recycled as separated electrical and electronic devices. Please observe the relevant
state-specific disposal regulations.

technical specification

Weight:
F.O. cable length:
Diameter of light spot at 50 cm distance: approx. 40 – 160 mm
Type of operation:
Ambient temperature:
Classification:
General Notes and Warnings
Check the correct operation of the product at regular intervals. Do not use the equipment if it shows visible
signs of damage.
- Use only light sources which are approved for use in a medical environment.
- Do not shine the light directly into the patient's eyes.
This product is a precise optical instrument. Please handle this product with care and store it in a clean place.
This product is not allowed to be entered or used in strong magnetic fields like MRI scanners.
HR and HRP binocular loupes.
®
i-View
®
ca. 230 g (cable supported by clips)
300 cm
continuous
+10°C to +40°C
moderate risk, IEC 62471 [2006]
med 3801
2013-12-16
7/28