Speed calibration
Speed calibration
Calibration fail example
● If a problem occurs such as the gear being extremely worn, correct data cannot be obtained at calibration and
"FAIL" is displayed� In this case, please request repair�
● Even if a malfunction does not occur, "FAIL" is displayed if the bur comes in contact with something during no-
load calibration or if you forget to apply a load at load calibration�
1
Remove the calibration bur from the calibration load�
2
Press
(CALIBRATION) with the bur not contacting
anywhere�
3
A beep sounds and the calibration bur automatically
rotates for a few seconds at low speed�
4
Calibration is complete when a beep sounds and
"PASS" appears on the display� Proceed to next step�
1
Press
(CALIBRATION) with the bur not contacting
anywhere�
2
A beep sounds and the calibration bur automatically
rotates for a few seconds at high speed�
3
Calibration is complete when a beep sounds and
"SUCCESS" appears on the display�
4
The LCD display will return to the normal screen�
19
Operation Procedure