11 – 2 Bluetooth specifications
Frequency band
Transmitting power
Modulation
Channels
Compatibility
11 – 3 Classification of equipment
● Type of protection against electric shock:
– Class l equipment
● Degree of protection against electric shock:
– Type B applied part
● Method of maintenance (cleaning) and sterilization recommended by the manufacturer:
– See "6 Post-use Maintenance"
● Degree of protection against ingress of water as detailed in IEC 60529 (EN 60529):
– Foot control: IPX8 (Protected against the effects of continuous immersion in water)
● Degree of safety of application in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide:
– Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide�
● Mode of operation:
– Non-continuous operation
11 – 4 Operation principle
Power is supplied to the control unit by operations on the foot control�
This makes the motor run and the bur attached to the handpiece rotate�
11 – 5 Warranty
NSK products are warranted against manufacturing errors and defects in materials� NSK reserves the right to analyze
and determine the cause of any problem� Warranty is voided should the product be not used correctly or for the
intended purpose or has been tampered with by unqualified personnel or has had non NSK parts installed� Replacement
parts are available for seven years beyond discontinuation of the model� Contact an authorized NSK dealer if repairs are
necessary�
11 – 6 Disposing of products
In order to avoid the health risks to operators handling the disposal of medical equipment, as well as the risks of
environmental contamination caused thereof, a surgeon or a dentist must to confirm the equipment is sterile�
Ask specialist firms who are licensed to dispose of specially controlled medical wastes to dispose the product for you�
2�4GHz ISM band (2�402-2�480GHz)
2�5mW[dBm]
GFSK
40channel 2MHz spacing
EN 300 328, EN 300 489-1, EN 301 489-17, EN 62479:2010
(Applied part: handpiece)