15 . 1 . ELECTROMAGNETIC COMPATIBILITY (EMC) INFORMATION

Medical electrical equipment requires special cautions regarding electromagnetic com-
patibility (EMC). Portable and mobile radio frequency (RF) communications equipment
can affect devices such as the Zen-O™ Portable Oxygen Concentrator. As such, the de-
vice should not be used adjacent to other equipment. If this is not practical, then observe
the device to make sure it is operating properly at all times.

15 . 1 . 1 . Guidance and manufacturer's declaration: electromagnetic emissions

The Zen-O™ Portable Oxygen Concentrator is intended for use in the electromagnetic en-
vironment specified below. The customer or the user of the concentrator should ensure
that it is used in such an environment.
EMISSION TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
COMPLI-
ELECTROMAGNETIC ENVIRONMENT/GUIDANCE
ANCE
Group 1
The Zen-O™ Portable Oxygen Concentrator uses RF
energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
Class B
The concentrator is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power
Class A
supply network that supplies buildings used for dome-
stic purposes.
Complies
35/338
EN