DECONTAMINATION (OPTION)
2
2
*If the H
O
decontamination is
2
2
terminated or ends in error, the
error number, error code, and error
message is displayed after
H
O
decontamination [Fig. 14].
2
2
You can see the error
information by pressing the
left/right keys.
*If the Cancel key (CANCEL) is
pressed to stop the decontamination
process during Step 1 or Step 2, a
confirmation screen is displayed on
the right side of the display [Fig. 15].
Change the setting to YES using the
up/down keys (△▽), and then press
the Enter key (ENTER).
Caution:
H
O
2
2
cannot be stopped during Step 3
(UV resolve).
If decontamination is canceled
during Step 1, the screen in Fig.
16 is displayed. Open the outer
door and inner door, remove the
power connector of the H
generator from the connector of
the main unit, and then remove
the H
O
generator to complete
2
2
the decontamination work.
If decontamination is canceled
during Step 2, the screen in Fig.
17 is displayed, Step 3: UV
resolve starts, and then the
screen in Fig. 18 is displayed
after the H
O
vapor has been
2
2
broken down. Open the outer
door and inner door, remove the
power connector of the H
generator from the connector of
the main unit, and then remove
the H
O
generator to complete
2
2
the decontamination work.
H
O
2
Door : Look
UV : ON
05/10
<Left side of the display>
H
2
Door : Look
<Left side of the display>
decontamination
H
2
Door : Closed
UV : OFF
<Left side of the display>
O
2
2
H
2
Door : Look
<Left side of the display>
H
2
Door : Closed
O
2
2
UV : OFF
<Left side of the display>
Decontamination
2
Step 1
Step 2
Step 3
Err35 :
O
Decontamination
2
Step 1
Step 2
UV : OFF
Step 3
O
Decontamination
2
Temp.
37.0
O
Decontamination
2
Step 1
Step 2
UV : ON
Step 3
O
Decontamination
2
Temp.
48.0
Step 3 : UV Resolve
90
min
Power Failure
<Right side of the display>
Abort?
NO
YES / NO
<Right side of the display>
Decon Aborted.
Please Remove
H
O
Unit.
2
2
<Right side of the display>
Step 3 : UV Resolve
90
min
<Right side of the display>
Decon Aborted.
Please Remove
H
O
Unit.
2
2
<Right side of the display>
[Fig. 14]
[Fig. 15]
[Fig. 16]
[Fig. 17]
[Fig. 18]
77