with a cloth saturated with multi-purpose cleaner and rinse with running water. Discard used cloth.
Dry with a clean cloth.
Disinfection:
To disinfect the storage caps, generously spray disinfectant solution on a clean cloth and wipe the
caps thoroughly. Discard used cloth. Allow disinfectant to air dry. Storage caps are not steam autoclave
sterilizable.
Disinfection of the sterilizable handpieces, handpiece adapters, reservoir bottles, blue interface caps
and sip tubes is not necessary prior to steam autoclave sterilization.
Packaging:
Place instrument in a paper or paper/plastic steam sterilization pouch. Instruments may be loaded into
dedicated instrument trays.
Sterilization:
(Handpieces, handpiece adapters, Reservoir Bottles, and blue interface caps with sip tubes)
Use a steam autoclave. Place bagged instruments into the steam autoclave, paper side up when
using a paper/plastic pouch. Disassemble reservoir bottles and interface caps before placing in pouch.
After warm-up is completed, operate at a sterilizing temperature of 273ºF/134ºC to 279ºF/137ºC and
pressure of 31psi/216kPa for 3 to 12 minutes, except when sterilizing the GRAY Steri-Mate handpiece,
the cycle time must be 4-12 minutes.
5