Ventilation Hose - Weinmann MEDUMAT Easy CPR Descripción Del Aparato E Instrucciones

Ocultar thumbs Ver también para MEDUMAT Easy CPR:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Lip
membrane
Disc
Disc
membrane
membrane
Expiration
Spontaneous
tube
breathing
tube

6.3 Ventilation hose

2. Dismantle the patient valve as shown in the ad-
3. Crinkled, misshapen and sticky lip and valve
4. Perform the hygienic preparation as described in
5. Reassemble the patient valve. When reassem-
6. Always perform a functional check before using
Caution!
Only reusable hose system WM 22520 (scope of
supply) is suitable for the hygienic preparation de-
scribed here. Do not subject disposable hose system
WM 28110 available as an accessory to hygiene
preparation. Replace it with a new one.
jacent diagram. It is neither necessary nor per-
missible to remove the membrane in the
spontaneous breathing tube for cleaning and
disinfection.
membranes must be replaced.
section 6.7.
bling, make sure that the lip membrane is cor-
rectly positioned.
the valve again (see "7.4 Checking patient hose
system" on page 50).
Hygiene preparation
EN