Technical Data; Device - Weinmann MEDUMAT Easy CPR Descripción Del Aparato E Instrucciones

Ocultar thumbs Ver también para MEDUMAT Easy CPR:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

11. Technical data

11.1 Device

Dimensions
WxHxD in mm
Weight incl.
accessories
Device category (93/
42/EEC)
Operation:
Temperature range
Humidity
Air pressure
Storage
Electromagnetic
compatibility (EMC)
according to
EN 60601-1-2
and EN 794-3:
– Radio interference
suppression
– Radio interference
resistance
Control unit
Gas supply
Gas for internal
control system
Operating pressure
EN
68
Technical data
MEDUMAT Easy CPR
100x145x90
incl. connections
approx. 0.6 kg
II b
–18 °C to +60 °C
max. 95 %
(1)
70 kPa
to 110 kPa
–40 °C to +70 °C
Test parameters and limit
values can be requested
from the manufacturer
(WEINMANN Emergency
Medical Technology
GmbH + Co. KG,
Frohboesestraße 12,
22525 Hamburg,
Germany).
EN 55011 B
EN 61000-4 Parts 2 to 6,
Part 11
timing pulse, volume
constant
Medical oxygen
0 l/min
(2)
2.7 to 6.0 bar
Minimum gas
volume required
Insp-exp. ratio (I:E)
Respiratory time ratio
(I:E) in CPR mode
Demandflow
Ventilation frequency
Tidal volume (V
)
t
V
tolerances:
t
Room temp. (20 °C)
–18 °C to +60 °C
Max. ventilation
pressure
O
concentration
2
Pressurized gas
connection
(1) Under normal atmospheric conditions, 70 kPa
correspond to a maximum altitude for use of
approx. 3,000 meters.
(2) 1 bar = ˆ 100 kPa
(3) 1 mbar = ˆ 1 hPa
MEDUMAT Easy CPR
70 l/min O
2
1:2 (V
≤ 150 ml),
t
1:3 (V
> 150 ml)
t
1:1
47-52 l/min
continuously variable
-1
from 10 to 25 min
continuously variable
from 65 to 950 ml
for ≤100 ml = ±20 %
for >100 ml = ±15%
for ≤100 ml = ±35 %
for >100 ml = ±20 %
**
or 30 ml
*
Tolerance related to
measured value
**
The larger tolerance
applies
(3)
20 or 45 mbar
100% O
(at least 98 %)
2
External thread G 3/8
*
*
*
*

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido