S S e e r r v v i i c c e e
R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r , , l l e e f f t t s s i i d d e e ) )
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
#2 Phillips screwdriver
#1 Phillips screwdriver
Diagonal pliers
Needle nose pliers
Straight pick
P P r r o o c c e e d d u u r r e e : :
1. Unplug the power supply from the S S e e c c u u r r e e C C o o n n n n e e c c t t .
2. Using a #1 Phillips screwdriver and needle nose pliers, disconnect the nurse call communication cable from the S S e e c c u u r r e e
C C o o n n n n e e c c t t .
3. Using a #2 Phillips screwdriver, remove the two screws (G) that secure the S S e e c c u u r r e e C C o o n n n n e e c c t t (F) to the mounting plate
(D) (Figure 3). Save the two screws.
4. Using a straight pick, remove and discard the label (521200380902) (C) that covers the O O N N / / O O F F F F switch (B) (Figure 2).
5. Turn the O O N N / / O O F F F F switch (B) to the O O F F F F position (Figure 2).
6. Using a #2 Phillips screwdriver, remove the ten screws (AC) that secure the front cover to the body (Figure 7). Save the
screws and front cover.
7. Using a #2 Phillips screwdriver, remove the two screws (AB) near the battery (V) that secure the cable routing tray (D) to
the body (C) (Figure 8). Save the screws.
EN
16
F F i i g g u u r r e e 7 7 – – R R e e m m o o v v i i n n g g t t h h e e c c o o v v e e r r