8. Using diagonal pliers, cut the cable tie (3000-300-114) (AK) that secures the 37-pin cable (M) to the cable routing tray
(D) (Figure 9). Discard the cable tie.
9. Move the 37-pin cable (M) and cable routing tray (D) up to remove the battery (700000341245) (V) from the body (C).
10. Unplug the two battery cables and discard the battery.
N N o o t t e e - - When you reinstall, make sure that the red wire connects to positive (+) and the black wire connects to negative
(-) on the battery.
11. Reverse steps to reinstall.
12. Verify proper operation before you return the product to service.