become so low as to be imperceptible.
The pressure indicated by the gauge needle at the moment the pulsations disappear is referred to as
the "diastolic or minimum pressure".
Diastole: minimum blood pressure level when the heart muscle is expanding and refilling with blood.
4) Now completely open the release valve until air begins to come out of the cuff. Blood pressure
measurement is now completed. To remove, pull the loose end of the cuff with a smooth motion.
SPECIFICATIONS
Pressure indication range:
Measuring Range of Cuff:
Accuracy:
Minimum scale value:
Operating Environment:
Specifications are subject to change without notice due to improvements in performance.

CLEANING, STORAGE AND MAINTENANCE

CLEANING
The Gauge, bulb and valve can be wiped with a slightly dampened cloth or alcohol pad. The Cuff can
be wiped with a slightly dampened cloth, or hand-washed with mild detergent and water solution (1:9
solution); in the second case, rinse and let air dry. Do not press or iron the cuff with a hot iron since to
the application of heat will melt the hook and loop and destroy their effectiveness
Please note:
- Do not allow liquids to enter inside the sphygmomanometer components.
- It is recommended a monthly cleaning but the following circumstances could request an increase of
the frequency:
- Before and after use of different patients;
- When the product is dirty
- High incidence of infectious diseases
- Contact with the skin when there is liquid, such as blood, medicine, sweat.
STORAGE AND MAINTENANCE
Always deflate the cuff completely before storage; needle should indicate zero when cuff is fully untouched
- fold the cuff with the hook and loop closure inside to protect it from dust or lint
- store the product at temperatures between -10ºC and 60ºC and relative humidity not exceeding 85%
- do not carry out self-repair or transformation
- do not drop or pull excessively on the sphygmomanometer components.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
0-300 mmHg
0-300 mmHg
≤ ± 3 mmHg or 2% whichever is greater
2mmHg
+10°C to +40°C,
15% - 85% relative humidity