Technical Data - Vaisala HUMICAP HMT130 Guía Rápida

Ocultar thumbs Ver también para HUMICAP HMT130:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 41

Technical Data

Table 2 Relative Humidity Measurement Specifications
Property
Measurement range
Accuracy (including non-linearity, hysteresis, and repeatability) at 0 ... +40 °C (+32 ... +104 °F):
0 ... 90 %RH
90 ... 100 %RH
Accuracy (including non-linearity, hysteresis, and repeatability) at -40 ... 0 °C, +40 ... +80 °C
(-40 ... 32 °F, 104 ... 176 °F):
0 ... 90 %RH
90 ... 100 %RH
Factory calibration uncertainty at +20 °C (+68 °F):
0 ... 90 %RH
90 ... 100 %RH
Humidity sensor
Table 3 Temperature Measurement Specifications
Property
Measurement range
Temperature sensor
Accuracy over temperature range:
At +15 ... +25 °C (59 ... 77 °F)
At 0 ... +15 °C and at +25 ... +40 °C
(at 32 ... 59 °F and at 77 ... 104 °F)
At -40 ... +0 °C and at +40 ... +80 °C
(at -40 ... 32 °F and at 104 ... 176 °F)
Table 4 Operating Environment Specifications
Property
Operating temperature of transmitter body, no
display
Operating temperature of transmitter body with
display
Operating temperature, HMP110 probe
Storage temperature
EMC compliance
14
Description/Value
0 ... 100 %RH
±1.5 %RH
±2.5 %RH
±3.0 %RH
±4.0 %RH
±1.1 %RH
±1.8 %RH
Vaisala HUMICAPâ 180R
Description/Value
-40 ... +80 °C (-40 ... 176 °F)
Pt1000 RTD Class F0.1 IEC 60751
±0.2 °C (±0.36 °F)
±0.25 °C (±0.45 °F)
±0.4 °C (±0.72 °F)
Description/Value
-40 ... +60 °C (-40 ... 140 °F)
-20 ... +60 °C (-4 ... 140 °F)
-40 ... +80 °C (-40 ... 176 °F)
-50 ... +70 °C (-58 ... 158 °F)
EN 61326-1 and EN 55022
M211281EN-D

Publicidad

Tabla de contenido
loading

Tabla de contenido