Replacing the optic illumination bulb
Taper Ring
Fig. 7
Figure shows X600WL
Bulb Holder
Bulb Holder
Bulb
Bulb
Joint
Joint
Fig. 8
14
<X WL Series>
<X BL Series>
(X WL, X BL Series)
Remove the handpiece from the coupling. Loosen and remove the
taper ring. (Fig. 7). Pull out the joint then remove the bulb holder,
then the bulb. Align the pins of the new bulb with the points in the
bulb holder and insert the bulb. Place the bulb holder back into the
joint. (Fig. 8). Firmly tighten the taper ring.
Optional Bulb : TA Bulb (Pack of 3) Order No.Y900132
Caution
Do not touch the glass section of new bulb. Screw the taper ring
tightly. A loose fitting taper ring could result in water and air
leakage. It is easier to screw in the taper ring when the bulb holder
is lightly pressed in by hand.
Replacing the O-rings
Replace the O-rings if water is present in the exhaust air line. This is an
Taper Ring
indication of possible water leakage within the coupling. ALWAYS
change the complete set of O-rings. Remove the handpiece from the
coupling. Loosen and remove the taper ring at the rear of the handpiece.
(Fig. 9) Gently remove each O-ring with your thumb. Insert the complete
Fig. 9
set of new O-rings in the correct grooves. (Fig. 10) Replace and firmly
tighten the taper ring.
O-rings
Optional O-ring : PTL O-ring Set (Pack of 5) Order No.Y900580
Make certain that the taper ring is firmly tightened.
Fig. 10
If the taper ring is loose water and air leakage could occur.
(X L, X Series)
Caution
15