Automatic lubrication
10
NSK Care3 Plus
This handpiece can be washed via Thermo Disinfector.
Lubricate the handpiece EACH TIME BEFORE AUTOCLAVING
Before lubrication, first clean the handpiece as previously described.
NSK Care3 Plus automatic handpiece lubrication system
After cleaning the handpiece simply connect the handpiece to the correct
adaptor and activate the Care3 System per the Care3 System
instructions.
Other brands of automatic handpiece lubrication systems
After cleaning the handpiece simply connect the handpiece to the correct
adaptor and activate the System per the manufacturer's instructions.
Manual lubrication
After cleaning the handpiece apply NSK PANA SPRAY Plus BEFORE autoclaving (Fig. 6). Make sure the CORRECT
application nozzle is connected to the plastic valve at the top of the spray can. Firmly insert the Nozzle into the rear of
the handpiece and activate the spray for 2-3 seconds until the spray exhausts from the handpiece head.
Nozzle
Caution
Always hold the NSK PANA SPRAY Plus can upright.
The spray lubricant is delivered from the can, into the handpiece, under
pressure. To prevent the handpiece from slipping from the Nozzle, always
hold the handpiece securily to the Nozzle.
Mount the arrow-head spray nozzle tip into the spray port and lubricate
chuck directly thorough the hole for bur attachment once a week.
Fig. 6
PANA SPRAY Plus