Integra Miltex Instrucciones De Funcionamiento página 3

Fuente de luz de lupas led
Ocultar thumbs Ver también para Miltex:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 6
3. Clip the power pack to your waist band or pocket.
a. Attach the lamp cord to the loupe frames with the small elastic strap.
b. Use the shirt clip on the lamp cord to position the cord on your shirt or lab coat as desired.
c. Remove battery from charger unit and slide into the power pack. The battery should snap into place.
4. Turn on lamp using on/off button.
a. The intensity control knob will adjust lamp intensity.
5. Adjust the direction of the LED beam.
a. To adjust the tension of the lamp head position, use the supplied Phillips screwdriver to tighten or
loosen the two small link screws as needed.
Operating Environment:
10° - 30°C (50° - 86°F)
95% or less relative humidity
70kPa – 106kPa
Transport/Storage:
0° - 40°C (32° - 104°F)
95% or less relative humidity
70kPa – 106kPa
Technical Data:
Headlight System: Internally powered by 7.2Vdc, 2500 mAH rechargeable Lithium-ion battery pack
Solaris (LED2-BP): 730mA
Power Supply: 100 – 240V~, 50 – 60 Hz, 0.5A
Warnings
Battery
Inspect batteries prior to use.
If the battery is dropped or receives a mechanical impact, ensure that the casing is not damaged.
Do not use damaged batteries.
The electrolyte contained within the lithium cells can cause irritation to the respiratory tract, eyes
and skin. In addition, cell venting could result in air contaminants, including corrosive or flammable
vapors. All precautions should be taken to limit exposure to the electrolyte vapor.
Light
Turn the light off when not in use. The lamp housing can get hot.
Do not use the LED lamp if ambient temperatures exceed 35°C (95°F).
When light is in use, always keep the front of the light at least 6 inches from skin and any surface
being illuminated.
Do not shine light into patient's eyes.
Do not shine light onto combustible material.
Do not allow contact of the lamp to the patient.
Use a curing filter when working with any light-curable resins.
Charging
When charging the battery, inspect the charging lights to ensure that charging is active.
Do not simultaneously touch the charging unit (when plugged in) and the patient.
Use only supplied charging station to charge the batteries. Li-ion batteries can be dangerous and
even explode if improperly charged. Batteries are designed to be charged only by the supplied
charging station.
Do not position the battery charger so it is difficult to disconnect the power adapter from the wall.
General
Check all components for burrs, nicks, and sharp edges prior to use.
If unit is jarred or dropped, re-inspect prior to use.
Do not use the system or charge the battery in an oxygen-rich environment.
Do not store liquids above the charger, power pack or any other component or accessory.
EN – ENGLISH
3

Publicidad

Tabla de contenido
loading

Tabla de contenido