• Please set the parameters according to the file manufacturer's recommendations.
• The use of torque reverse function can effectively protect the file from separation within the canal.
• If the auto reverse is too frequent during use, please recapitulate, irrigate and lubricate the root
canal or increase the torque as per file manufacturer's recommendations.
9. 4. Additional settings (Fig. 9.G)
1 - When the device is powered off, press simultaneously the S and the main buttons.
2 - Touch S to select the advanced parameter to be set.
3 - Touch < / > (- / +) to set values.
4 - Press the main switch to escape this mode and come back to the standby screen.
Volume of beep (BEEP VOLUME)
9.G.1
Low (Low), middle (Mid) or high (H igh) volume of
audible signals. Factory setting: Middle (Mid).
Auto Power off (AUTO P.W.R)
9.G.2
Automatic power off after 3, 5, 10 or 15 min.
Factory setting: 10 min.
Orientation of the display (Hand)
9.G.3
Right (RIGHT) or left (LEFT) orientation.
Factory setting: RIGHT Hand.
When connected to Dual Pex apex locator, Dual Move offers advanced functions, see section 10 to 13.
9. 5. Calibration (Fig. 9.H)
Connect the contra angle to the Dual Move handpiece. Do not insert a file.
Enter the calibration interface (see section 9.4) then choose 'ON'.
9.H.1
Press the main switch to start the calibration.
9. 6. Error codes
The battery is flat, auto power off will occur. Charge
9.I.1
the device immediately.
9.I.2
Motor overload. Reduce the load.
• Follow the instructions of the file manufacturer to set the functions and parameters.
• Calibration is strongly recommended after 10-20 autoclaving cycles for the contra-angle and after
its replacement.
• Calibration shall be performed with a 50% battery level as a minimum.
• Do not apply pressure to the contra-angle during calibration.
• If an error code is displayed, please contact your after-sales service.
connected to
Refer to the Dual Pex user manual to install it correctly before connecting to Dual Move.
10. Dual Pex to Dual Move connection (Fig. 10)
1 - Make sure that Dual Move is in standby mode.
6
Inactivity time (SET TIME)
9.G.4
Automatic return to the standby display after the
specified time of inactivity: 1-15s. Factory setting: 10s.
Calibration (Calibration)
9.G.5
Automatic calibration of speed and torque. See next
section.
Bars show the progress.
9.H.2
2 beeps when completed.
Long use time, risk of overheating. Stop using the device
9.I.3
for a while.
(apex locator)
2 - Check that the protective silicon sleeve correctly covers all the metallic surface of the contra-angle.
3 - Open the rubber cover and plug the data transfer cable to Dual Move.
4 - Power ON Dual Pex and connect the data transfer cable to Dual Pex. CONNECTED ! is displayed on the Dual Move's
screen when the connection is correctly done.
11. Connection testing (Fig. 11)
Fig. 11.A Insert a file into the contra-angle, create a short circuit between the lip clip and the file.
The file may start rotating if "Auto start / stop" function (section 13, Fig. 13.C). is activated.
Fig. 11.B Press the Dual Move power switch. A full-length white bar appears next to the line representing the canal, indicating that the
system is working properly and is ready to be used.
Hang the lip hook on the patient's lower lip and start the treatment.
12. Dual Move screen during treatment (Fig. 12)
Fig. 12
The white strip on the Dual Move screen shows the progression of the file into the canal.
The beeps frequency increases when the file approaches the apex.
13. Combination functions (Fig. 13)
The following advanced functions are only usable when Dual Move is connected to Dual Pex.
1 - Touch < / > to select a memory number (M1...M6)
2 - Touch the S key for more than one second to enter the advanced function setting interface.
3 - Touch < / > to activate (ON) or deactivate (OFF) the advanced functions. If "CONNECT FUNCTION" is deactivated
(OFF), all the advanced functions will be inactive, even if they are individually activated (respective status set to "ON").
The code of following function to be set (APA, ASS, ASD) is displayed at the bottom of the screen, just next to the
battery status.
4 - Touch S to switch to the following combination function.
5 - Touch < / > to activate (ON) or deactivate (OFF) the function then S to go to the next function.
Automatic return to the standby screen if no action for a while (time-out to be set, factory value 5s).
Connection functions (CONNECT Function)
When Dual Move and Dual Pex are connected, set "ON"
13.A
to have the possibility to use the combination functions,
depending on their status.
Apical auto-reverse (APA)
Automatic reverse when the file reaches the reference
13.B
point (*). Automatic restart as soon as it comes back
above the reference point.
Factory status of advanced function for each Memory:
Function
M1, M2, M3
Apical Reverse (APR)
Activated (ON)
Auto Start & Stop (ASS)
Activated (ON)
Apical Slow Down (ASD)
Deactivated (OFF)
(*) Position of the Reference point (RP) is set through the Dual Pex interface then automatically sent to Dual Move and
displayed as a cursor on the Dual Move screen. When the file reaches the reference point, Dual Move will start the
APICAL Reverse, APICAL Slow Down depending on their respective status (ON / OFF).
Auto Start stop (Auto Stt Stp - ASS)
When the file enters the orifice of the canal, the motor will
13.C
automatically start. When the file leaves the orifice of the
canal, it will automatically stop.
APICAL Slow Down (ASD)
Automatic deceleration once the file exceeds the reference
13.D
point (*).
M4, M5, M6
Deactivated (OFF)
Activated (ON)
Activated (ON)
7