5. Connection of the Spray Nozzle (FX23/FX23m/FX25/FX25m)
1) Attach the Spray Nozzle (Optional) to the handpiece head.
2) Cut the Irrigation Tube to the desired length then connect to the Spray
Tube Nozzle on the motor. (Fig.4)

6. Check before treatment

Check that the Head Cap is firmly tightened. Also check for handpiece
vibration, noise and overheating. If any abnormalities are found do not
use the handpiece and contact your Authorized NSK Dealer.
7. Maintenance
After each patient maintain the product as follows.
7-1 Cleaning (Handpiece)
1) Remove dirt and debris from the product. Do not use a wire brush.
2) Wipe clean with alcohol-immersed cotton swab or cloth.
This icon denotes that the product can be washed via Thermo
Disinfector. (FX15m/FX23m/FX25m/FX75m)
4
Refer to the Thermo-Disinfector manual.
CAUTION
After washing with Thermo-Disinfector and prior to lubrication,
dry the product until all internal moisture is thoroughly removed.
Thermo-Disinfector moisture remaining inside the product could
reduce the effect of lubrication and could cause corrosion inside
of the product.
To clean the product never use any solvent such as benzine or
thinner.
7-2 Lubrication
NSK PANA SPRAY Plus
Apply NSK PANA SPRAY Plus every time after each use and/or before
autoclaving.
1) Remove the bur or file from the handpiece.
2) Insert the Spray Nozzle into the Spray Port nozzle on the can.
3) Insert the Spray Nozzle in rear of the handpiece. Hold the
handpiece and spray for approximately 2-3 seconds. Apply