lubricant until it expels from the handpiece head for at least 2
seconds. (Fig.5)
CAUTION
When applying spray be sure to hold the handpiece firmly to
prevent the handpiece from slipping out of the hand due to the
spray pressure.
Hold the spray can upright.
NSK Care3 Plus
When using NSK Care3 Plus automatic handpiece cleaning and
lubrication system, refer to the Care3 System instructions.
7-3 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur or file
after each patient and sterilize as noted below.
1) Insert an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or
3 min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required
for use.
CAUTION
Do not autoclave the product with other instruments even when it
is in a pouch. This is to prevent possible discoloration and damage
to the product from chemical residue on other instruments.
Keep the product in suitable atmospheric pressure, temperature,
humidity, ventilation, and sunlight. The air should be free from
dust, salt and sulphur.
Immediately after use, the handpiece should be cleaned,
lubricated and sterilized. If blood remains on the external or
internal surfaces it can become clotted and cause rust.
Do not heat or cool the product too quickly. Rapid change in
temperature could cause damage to the product.
If the sterilizer chamber temperature may exceed 135°C during
the dry cycle then delete the dry cycle.
5